2.2.26 General
1. Storing trend data
Check that the monitor is capable of storing the trend information and temporary
settings in a short (max. 15 minutes) standby situation with no power cord.
"
2. Service reset
Check the Service Reset switch. Switch the monitor to standby and press the Service
Reset switch for at least five seconds. Switch the monitor back on and check that the
monitor performs a Cold Start, i.e. all trend information is cleared.
"
3. Watchdog
Enter the Set/Test service menu and perform Watchdog
Check that the monitor restarts.
"
4. Service Log reset
Enter the Service Log service menu.
Clear the content of the Service Log by selecting Reset Log from the menu.
"
5. Electrical safety check
Perform an Electrical safety check and a leakage current test. Check that the monitor and
all connected units function normally after the performed test.
"
6. Final cleaning
Switch the monitor to standby and perform final cleaning.
Fill in all necessary documents.
"