Document no. M1144955-001
Record Data: A selection to print out the shown
service data to the device defined in the ...Service -
Record Data menu.
TONO
All the numbers in this section are AD counts, i.e.
computer internal data.
EXTERNAL ADC
CO2; CO
2
sensor CO
2
signal.
temp; CO
2
sensor temperature signal.
cath press; pressure measured from the catheter
line.
pump press; pressure measured from the pumping
unit.
INTERNAL ADC
VDD; 5 V digital units driving voltage received from
the module frame.
5V; 5 Volts of the module derived internally from
15vdc.
15V; 15 Volts of the module received from the module frame.
pneuma voltage; 15 V voltage driving pneumatic elements of the module, derived from 15 V dirty.
CO2 det offset; CO
2
detector offset voltage.
lamp current; CO
2
sensor lamp current.
15V dirty; 15 V voltage received from the module frame.