Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
22
Document no. M1144956-003
4. Net Unit and Synchronization Unit
Check that:
• screws are tightened properly
• the connector pins are intact
"
5. Front Unit
Check that:
• the Connector Board screws are tightened properly and the cable connectors are
clean and intact
• the Inverter Board screws are tightened properly and the cable connectors are
clean and intact
• the Backlight cables are intact and attached properly
• the keyboard foil cables are intact and attached properly to connectors on the
Connector Board
• the Connector Board foil cable is intact and attached properly to the connector on
the Central Processing Board
• the video signal foil cable is intact and attached properly to its connector on the
Central Processing Board
Reattach the Front Unit.
"
6. External parts
Check external parts of the Display Unit:
• all stickers are intact
• the LCD display screen is intact
• the ComWheel cover is intact and attached properly
• all mechanical parts are functioning properly
Connect the Display Unit to the Frame Unit.
"
7. Inserting module
Check that the module used goes in smoothly and locks up properly in both module slots.
Leave the module connected.
"
8. Fan filter
Clean or replace the Display Unit fan filter.
"