Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
22
Document no. M1144956-003
4. Net Unit and Synchronization Unit
Check that:
screws are tightened properly
the connector pins are intact
"
5. Front Unit
Check that:
the Connector Board screws are tightened properly and the cable connectors are
clean and intact
the Inverter Board screws are tightened properly and the cable connectors are
clean and intact
the Backlight cables are intact and attached properly
the keyboard foil cables are intact and attached properly to connectors on the
Connector Board
the Connector Board foil cable is intact and attached properly to the connector on
the Central Processing Board
the video signal foil cable is intact and attached properly to its connector on the
Central Processing Board
Reattach the Front Unit.
"
6. External parts
Check external parts of the Display Unit:
all stickers are intact
the LCD display screen is intact
the ComWheel cover is intact and attached properly
all mechanical parts are functioning properly
Connect the Display Unit to the Frame Unit.
"
7. Inserting module
Check that the module used goes in smoothly and locks up properly in both module slots.
Leave the module connected.
"
8. Fan filter
Clean or replace the Display Unit fan filter.
"