Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
24
Document no. M1144956-003
14. Loudspeaker
Check the loudspeaker by setting the alarm sound:
Alarms Setup
Alarm Volume
Test the whole volume scale from 1 to 10 by turning the ComWheel and check that the
alarm volume changes correspondingly. The alarm sound should be clear and audible
with all the settings.
"
15. Display brightness
Check the display brightness control:
Monitor Setup
Display Brightness
Press the ComWheel and check that the display brightness follows the selected
brightness.
"
16. Monitor software
Enter the service menu:
Monitor Setup
Install Service (Password 16-4-34)
Service (Password 26-23-8)
Take down the information regarding Monitor and Power Management Board software.
"
17. Content of service log
Select Service Log from the menu.
Check the contents of the Service Log for possible problems. If the monitor contains the
recorder unit, record the Service Log on the recorder by selecting Record Log.
"
18. Power supply output voltages
Check the power supply unit output voltages through the service menu:
Monitor Setup
Install Service (Password 16-4-34)
Service (Password 26-23-8)
Frame
Power Supply
The Display Unit voltages should meet the following ranges:
+12V 11.50...12.50 V
+3.3 V 3.00...3.60 V
Vbus 10.00…17.00V
Depends on the power source