Frame Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
25
Document no. M1144958-001
4.2 Recorder Unit
Problem Cause What to do
Recorder will not start.
No error messages shown.
Connector board loose. Check connector board connections.
Recorder board faulty. Replace the recorder board.
Recorder unit faulty. Replace the recorder unit.
Recorder works but nothing is
printed on the paper.
Active side of the paper is
downwards.
Turn the paper roll over.
To test which side is active: Place the
paper on a hard surface and draw a
line with a fingernail - a dark line will
appear on the active (thermal) side.
Recorder unit faulty. Replace the recorder unit.
Message Explanation
Recorder: out of paper Release paper jam or insert a roll of paper into the
recorder.
Recorder: cover open Close the recorder cover properly.
Recorder: thermal array overheat Recorder overheated. Stop using and allow it to cool
down.
Recorder: input voltage low +15 Vrec is too low. Check the connectors and
connector board.
Recorder: input voltage high +15 Vrec is too high. Check the connectors and
connector board.
Recorder system error
1, 2, 3
System error.
If the problem persists, replace the recorder unit.
Recorder: module removed A communication error due to a fault in the recorder
or in the Central Processing Board.