GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 4 - Functional Checks 4-17
4-4-5 Volume Mode Checks
NOTE: Different menus are displayed depending on which Menu and which Volume Mode
(Static 3D, Real Time 4D, 4D Biopsy VCI A-Plane*, VCI C-Plane*, STIC*, TUI*) is selected.
4-4-5-1 Pre-Volume Mode Functions
NOTICE
Real Time 4D (freeware on Voluson® 730ProV), RT_4D_Biopsy*, VCI*, STIC*, TUI* and VOCAL II* are
Options. If these options is not a part of the system configuration, the appendant checks can be omitted.
BT Version:
The VCI*, STIC* and TUI* Options are only available at systems upgraded to BT´05 (SW 5.x.x).
Figure 4-12 Pre-Volume Mode menus
Table 4-7 Pre-Volume Mode Functions
Step Task Expected Results
1
3D RENDERING
3D Static volume acquisition + rendered 3D image (also in combination with PD or CFM Mode)
1
SECT. PLANES
3D Static volume acquisition resp. 4D volume acquisition without rendered 3D image
Volume Acquisition: Static 3D
4. Start the Volume Acquisition with the Freeze
key resp. the right trackball key.
Volume Acquisition: Real Time 4D
1. select the desired Volume Acquisition Mode:
The Volume Mode function is switched on, the “3D Pre” respectively “4D Pre” menu appears
on the screen (write mode) and the volume box appears on the Image area.
2. select Acquisition and/or
3. select an User Program
Visualization Mode