GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 4 - Functional Checks 4-33
4-5-2 Load User Settings Only (Application Settings)
1.) Insert the DVD/CD+(R)W or MO (Magneto-Optical Disk) into the drive.
2.) Press the UTILITIES
key on the control panel. The menu area changes to the Utilities menu.
3.) Select the SYSTEM
item from the menu area to activate the setup desktop screen.
4.) Select the
BACKUP page and click the LOAD button of the “User Settings Only” group.
5.) Choose the media (e.g., DVD/CD+RW) and click the LOAD
key.
6.) Select the appropriate file and click OK
. The “Load Backup Data” window appears.
CAUTION
The loading procedure overwrites the existing application settings on the local hard drive. Make sure to
insert the correct DVD/CD or MO. Additionally you can load the backup from D:\User Settings.
Figure 4-24 User Settings Only - Load window
NOTICE
It is highly recommend to use Application settings which are adapted for the systems software version!
Figure 4-25 Load Backup Data
Choose the media
and
then click Load
>> button
Load
button
Complete Backup