EasyManua.ls Logo

GE Voluson 730Pro - Software and Functional Checks after the Upgrade; System Setup - System Info Page

GE Voluson 730Pro
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
8-14 Section 8-7 - Software and Functional Checks after the Upgrade
Section 8-7
Software and Functional Checks after the Upgrade
1.) Press the UTILITIES key on the control panel once to display the Utilities menu.
2.) Select the SYSTEM
item from the Utilities menu - which is displayed on the left side of the screen.
3.) Select
SYSTEM INFO to see which Software/Hardware version is installed in the unit.
4.) Verify the correct settings of the
OPTION page; see: Figure 8-2 on page 8-3.
If necessary, customize the settings according to the printout.
5.) Check and (if necessary) match the
NETWORK page settings according to the printout:
• “DICOM Configuration” dialog page; see: Figure 8-3 on page 8-4
• “Network Configuration” dialog page; see: Figure 8-4 on page 8-4
6.) Restart the system and perform basic functional checks to ensure system is functioning normally.
Figure 8-20 System Setup - System Info page
Serial number
Application Software
Move the scroll bar downwards to review additional information about installed software/hardware.
Serviceplatform
information
Backup File Name
System Type
Date of Factory Settings
Hardware ID
EUM Version
Operating System
(e.g., Service Pack 4)
SLOT_CPU
(e.g., KONTRON)

Table of Contents