GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-55
1.) Click the MAINTENANCE REPORT button. The following message box will be displayed.
2.) Fill in the requested information and click OK
.
3.) Click the EXIT
button on the Service Tools window and the EXIT button on the System Setup
Service page.
NOTE: After Hardware or Software modifications normally the “Maintenance Report” message box
(see: Figure 5-41) appears automatically on the screen.
5-15-3-7 Monitor Test
1.) To perform the Monitor test, select the MONITOR TEST
button.
The following message appears on the screen.
2.) Read the displayed instructions. Afterwards confirm with YES
.
Figure 5-41 Maintenance Report
Figure 5-42 Monitor Test - Instructions