3-8-1-1 Physical Dimensions of Voluson® 730Pro / 730ProV
The physical dimensions of the Voluson® 730Pro / 730ProV unit are summarized in Ta ble 3 - 5.
Table 3-6 lists the size of Voluson® 730Pro / 730ProV, with monitor and without on-board peripherals.
3-8-1-2 Weight without Monitor and Peripherals
3-8-1-3 Acoustic Noise Output
max. 57dB(A)
3-8-2 Electrical Specifications
Power Consumption nominal 1010 VA including all options.
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories.
All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
Output voltage for: ST1 - ST5: 150V or 230V.
Table 3-5 Physical Dimensions of Voluson® 730Pro / 730ProV
Height Width Depth
142 cm / 55.9 inches 68 cm / 26.8 inches 100 cm / 39.4 inches
Table 3-6 Weight of Voluson® 730Pro / 730ProV with Monitor and without other Peripherals
Model Weight [kg] Weight [lbs.]
Voluson® 730Pro / 730ProV 136 300
Table 3-7 Electrical Specifications for Voluson® 730Pro / 730ProV
Voltage Tolerances Current Frequency
100 VAC ±10% 10.10 A 50, 60 Hz (±2%)
110 VAC ±10% 9.20 A 50, 60 Hz (±2%)
115 VAC ±10% 8.80 A 50, 60 Hz (±2%)
130 VAC ±10% 7.80 A 50, 60 Hz (±2%)
230 VAC ±10% 4.40 A 50, 60 Hz (±2%)
240 VAC ±10% 4.20 A 50, 60 Hz (±2%)
CAUTION
Modification of voltage setting only by an authorized service person!
The maximum power consumption of equipment (inclusive color video monitor) connected to
these outlets must not exceed 350VA !