GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
3-18 Section 3-5 - Connection of Auxiliary Devices
3-5-10 External USB-Devices
3-5-10-1 Connection
When an external USB-storage device (such as a USB-memory stick or an external hard disk) is
connected to the Voluson® 730Pro / 730ProV system, Windows detects the device and automatically
installs a driver.
During this process, several dialogs may pop up, starting with the „Found New Hardware“ dialog.
The device is then accessible using the drive letter the system assigned to it.
3-5-10-2 Disconnection
Before an external USB-device (e.g., USB-memory stick) can be disconnected, the system has to be
informed about the removal of the device! For this purpose the System Setup -
BACKUP page (see:
Figure 4-22 on page 4-31) has a STOP USB DEVICES
button.
NOTE: Unplugging or ejecting devices without first stopping them can often cause to crash and lose of
valuable data.
By clicking the STOP USB DEVICES
button, the „Unplug or Eject Hardware“ dialog is started.
Using this dialog, the USB-devices can be stopped before they are physically disconnected.
The „Unplug or Eject Hardware“ dialog shows all USB-devices that are connected to the system.
On every system is an USB mass storage device, the DVD/CD writer, which has the drive letter (F:).
If the system has an optional MO-drive installed as well, it is listed too, and has the drive letter (E:).
Figure 3-18 Found New Hardware - USB Device
NOTICE
When connecting external USB devices, be sure to execute Safety Directions found in Chapter 2 of the
Voluson® 730Pro / 730ProV Basic User Manual.
Figure 3-19 Unplug or Eject Hardware