GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-5
8-2-5 Software - Installation Procedure (via Service Page)
1.) Insert the System DVD into the drive.
2.) Press the UTILITIES
key and then select SYSTEM item from the Utilities menu.
3.) Select the
SERVICE page.
4.) Type in the password SHE and click ACCEPT
.
5.) Click on the UIS
button for updating the Ultrasound Application Software.
Figure 8-5 System Setup Service
Figure 8-6 Service Tools
UIS button