GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-43
3-8-4 On-Board Optional Peripherals
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories.
All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
Output voltage for: ST1 - ST5: 115V or 230V.
CAUTION
Modification of voltage setting only by an authorized service person!
The maximum power consumption of equipment (inclusive color video monitor) connected to
these outlets must not exceed 350VA !
Table 3-8 Approved Peripherals
Device Manufacturer Model Video Signal
B/W Video Printer SONY UP-895MD NTSC/PAL
Digital Color Video Printer SONY UP-D21MD USB-Port
Digital Color Video Printer SONY UP-D23MD USB-Port
Video Cassette Recorder
SONY
SVO-9500MD
SVO-9500-MDP
NTSC
PAL
Mitsubishi
HS-MD3000U
HS-MD3000E
NTSC
PAL
Line Printer Hewlett Packard hp deskjet 990cxi
Line Printer Hewlett Packard hp deskjet 995c
Bluetooth Printer
Hewlett Packard HP 5600 Series
Olivetti Olivetti Job Jet 210
Magneto-Optical Drive (MOD) Fujitsu MCM3130AP
Global Modem Multi-Tech MT5634ZBA
ECG Preamplifier MAN6
Footswitch MFT7