GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-21
3-6-2 Power Off/ Shutdown
NOTE: After turning off a system, wait at least ten seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
3-6-2-1 Back-end Processor Power Down
1.) Press the ON/OFF Standby switch left below the Control Panel; see: Figure 3-23.
3-6-2-2 Scanner Shutdown
1.) Press the ON/OFF Standby switch left below the Control Panel.
2.) Switch OFF the Circuit Breaker at the rear of the system.
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
3.) Disconnection of the Mains Power Cable is necessary. For example: Repairing the scanner.
Unscrew the 2 screws and remove the pull-out protection to disconnect the cable from the system,
or unplug the cable from the wall socket.
4.) Press once on the brakes to block the front wheels (brakes on front wheels under the foot rest).
5.) Prior to disconnect a probe freeze the image.
6.) Open the right-hand side door, remove the probe cable from the cable holder and close the door.
7.) Turn the probe locking handle counterclockwise. Pull the connector straight out of the probe port.
Figure 3-24 Circuit Breaker, Protection and Power Cable on back of Voluson® 730Pro / 730ProV
CAUTION
If a probe is disconnected while running (Write-Mode) a software error may occur.
In this case switch the unit OFF (perform a reset).
Circuit Breaker
Mains Power Cable
pull-out Protection
2 screws