GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
4-44 Section 4-9 - Site Log
Section 4-9
Site Log
4-9-1 Site Log - System (Service Database)
1.) Press the UTILITIES key on the Control Panel. The menu area changes to the Utilities menu.
2.) Select the SYSTEM
item from the menu area to activate the setup desktop screen.
3.) Select the
SERVICE page. The “password window” appears automatically.
4.) Enter the password SHE and click the ACCEPT
button to display the Service Tools window.
5.) Click the MAINTENANCE REPORT
button. The following message box will be displayed.
6.) Fill in the requested information and click OK
.
7.) Click the EXIT
button on the Service Tools window and the EXIT button on the System Setup
Service page.
NOTE: After Hardware or Software modifications normally the “Maintenance Report” message box
(Figure 4-40) appears automatically on the screen.
Figure 4-39 System Setup Service page and Service Tools window
Figure 4-40 Maintenance Report