Service Platform (SSW) Upgrade Procedure
8-5-1 Manpower
One Person, 10 min.
8-5-2 Tools
System DVD or Upgrade CD
8-5-3 Upgrade Procedure
1.) Insert the System DVD or Upgrade CD into the drive.
2.) Press the UTILITIES
key on the control panel once to display the Utilities menu.
3.) Select the SYSTEM
item from the Utilities menu - which is displayed on the left side of the screen.
4.) Select the
SERVICE page.
5.) Enter the password SHE, and click the ACCEPT
.
6.) The “Service Tools” menu appears on the screen.
7.) Click on the SSW
button for updating the Service Platform Software.
Figure 8-11 Service Tools