GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-13
3-5-5-2 Sony SVO-9500MD
Figure 3-12 Video Recorder Connection Scheme
Figure 3-13 DIP Switches
NOTICE
Please use the proper connection set and remote cable.
see: Section 9-12 "Optional Peripherals and Accessories" on page 9-25.
Set DIP Switches on back of the VCR as shown:
Switch 1: OFF-Audio is muted during fast playback
Switch 2: OFF-FF/REW mode is used during searches.
Switch 3: ON-VCR counter is reset upon cassette ejection.
Switch 4: OFF-The VCR can be controlled by the system.
Switch 5 + 6: ON-The Baud rate is set to 19200 bit/sec
(must be the same as Voluson® 730Pro / 730ProV)