GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
7-20 Section 7-7 - Troubleshooting Trees and Instructions
7-7-7 Printer Troubleshooting (cont’d)
Figure 7-21 Printer Troubleshooting (cont’d)
Still unable to use the Printer(s)!
Check the signal cables between
GEF and Backpanel.
YES
NO
Is printing unavailable
even by manual
operation?
Still no image on the print
out
Video Output of System o.k.?
NO
YES
Printer o.k.
Replace the Printer
Replace CKV
NO
YES