GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-25
CtrlMvGrpRepresentations::isInWriteMode not supported for
mode nr xx
restart the system
CW quant TX Frequ out of range restart the system
CW_BaseLinePos Out Of Range press ok and save this user-setting once again
CW_Gain Out Of Range press ok and save this user-setting once again
CW-hardware doesn't support pencil probe + CW-Hardware not available->HW problem
CW-HW-PRF == 0 restart the system
CW-TxFrequency == 0 restart the system
CW-TxPower Out Of Range press ok and save this user-setting once again
D:\\SERVICE\\ConfIndexFile is corrupt restart the system
Datatype misalignment restart the system
Delete error (Backup data could not be deleted.)
check storage destination of Full Backup
(e.g., DVD/CD, insufficient rights on target Network drive,
write protection on MO or USB-drive)
Density Out Of Range press ok and save this user-setting once again
Different software error (Backup data was made on another
system with a different software version.)
This backup cannot be restored by the user.
Disc full! use a new CDR for writing data to CD
Display:Rect Region fails restart the system
Division by zero restart the system
done is low! restart the system
Downgrade error (Backup data was made with a software
version higher than the installed version.)
load appropriate backup for installed version
DSP - Gamma Corr - Load Data Timeout restart the system
DSP - HilbertCoeff - Load Data Timeout restart the system
DSP - Low Pass Coeff - Load Data Timeout restart the system
DSP - SetFFT_Para - Load Data Timeout restart the system
DSP - SetWMF_Koeff - Load Data Timeout restart the system
DSP/MSE:Hanning-Window Load Data Timeout restart the system
ECG_InterfaceToHW::vLookupForNewSamples() allocating
expanded buffer[xxx] for TimeStamps failed!!
restart the system
Electronic user manual not installed. Please install. install Electronic User Manual (EUM) and try it again
End Bandwidth too big restart the system
End ET too big restart the system
End frequency too big restart the system
Error Messages Actions