GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-7
8-2-5 Software - Installation Procedure (via Service Page) (cont’d)
9.) Complete the Software installation with the FINISH
key.
The system restarts automatically.
10.)If desired, load the appropriate factory settings; see Section 8-3 on page 8
or the full backup; see: Section 8-4 on page 8-8.
11.)If necessary, update the Service Software (SSW) as described in Section 8-5 on page 8-9.
12.)Afterwards update the Electronic User Manual (EUM) as described in Section 8-6 on page 8-11.
13.)Perform Software and Functional checks as described in Section 8-7 on page 8-14.
Figure 8-10 V730 Software Setup