QUICK START GUIDE
Setting Method Parameters
To display the method parameters, press . The
View/Modify Method
screen will be
displayed.
Only certain parameters can be changed.
For this titration, the NaOH titrant concentration and the size of the HCl sample need to be
entered.
To accomplish this:
• Highlight
Titrant Conc.
option, then press .
The
Titrant Concentration
screen will be displayed.
• Enter the correct value, then press .
• Highlight
Analyte Size
option, then press .
• Enter the volume of the sample (ex: 5 mL), then
press .
• Press , highlight
Save Method
option and then
press .
Setup Titration Report
Users can select the information that is stored for each titration.
To setup the titration report, follow the procedure below:
• From the main screen, press . The
Data Parameters
screen will be dis-
played.
• Highlight
Setup Titration Report
and press .
• Mark the fields to be included in the titration report with the “*” symbol. Use the
and keys to highlight a field and / to toggle the field.
• Press to save the customized report .