7-28
pH, mV & ISE MODE
7.3.2.15 Stirrer Configuration
Set the stirrer configuration: Stirrer 1, Stirrer 2 or Disabled.
7.3.2.16 Stirring Speed
The stirring speed for the selected stirrer can be set.
7.3.3 ISE Calibration
It is recommended to calibrate the instruments frequently if high accuracy is required. The
instrument should also be recalibrated whenever the “Calibrate Electrode” message appears
on the LCD.
Due to electrode conditioning time, the electrode must be immersed for several seconds to
stabilize. The user will be guided step by step during calibration with easy-to-follow mes-
sages on the display. This will make the calibration a simple and error-free procedure.
PREPARATION:
Pour small quantities of the standard solution into clean beakers. If possible, use plastic
beakers to minimize any EMC interferences.