4-8
GENERAL OPTIONS
4.6 Language
Select an available language.
4.7 Total Volume Alert
This screen allows a programmable reminder to appear when the titrant reservoir is below
100 mL. The titrant volume will decrease as the titrant is used.
The “Low Titrant Volume” reminder message will appear when the available titrant volume
is under 100 mL.
After the new titrant volume has been set on the titrator (in the
Total Volume Alert
screen),
a warning message appears reminding the user to perform titrant re-standardization. The
volume of titrant can be set from 0 to 10,000 mL.