QUICK START GUIDE
SAFETY MEASURES
The following safety measures must be followed:
1. Always ensure that the power supply cable is connected to a grounded main power plug.
2. Never connect or disconnect the pump assembly with the titrator turned on.
3. Verify that the burette and the attached tubing are assembled correctly.
4. Always check that the titrant bottle and the titration beaker are placed on a flat, stable
surface.
5. Always wipe up spills and splashes immediately.
6. Avoid the following environmental working conditions:
• Severe vibrations
• Direct sunlight
• Atmospheric relative humidity above 95% non-condensing
• Environment temperatures below 10
°
C and above 40
°
C
• Explosion hazards
7. Have the titrator serviced by qualified service personnel only.