5-13
METHODS
5.5.6 End Point Mode
5.5.6.1 Fixed End Point (pH or mv)
Fixed End Point (pH):
The titration is terminated when the preset pH value has been exceeded. The end point
volume is a calculated value based on the dispensed volume when pH is under the preset
value and the dispensed volume when pH exceeded the preset value.
The range is from - 2.000 to 20.000 pH.