7-23
pH, mV & ISE MODE
• Press or to add/remove the selected standard to/from standard Group.
• Press to return to ISE Setup menu.
7.3.2.6. Calibration Reminder
In order to have accurate readings, the electrode must be calibrated frequently. Three op-
tions are available for the calibration reminder:
Daily
- the calibration reminder will appear daily at specified time.
Periodic -
the calibration reminder will appear after the set time has elapsed since the lost
calibration.
Disable
- the calibration reminder will not appear.
7.3.2.7 Set Reminder Period
If Daily or Periodic option was selected for the Calibration Reminder, the reminder period
must also be set.
For a daily reminder period the time of day can be set.
For a periodic reminder period the number of days, hours and minutes can be set.