Loading...
5-27
METHODS
5.5.13.3 Sample Calculations by Weight
This calculation is used when the concentration of an analyte is determined by the weight of
the sample. The results are based on the initial sample weight (in grams). When you choose
this formula, select the
Titrant Unit
first and then the
Final Result Unit
.
The titrator will calculate the results based on the selected units.
The titrator will provide the results based on the titrant and sample units selected.
Titrant Units:
M (mol/L) moles/liter
N (eq/L) equivalents/liter
g/L grams/liter
mg/L milligrams/liter
Final Result Units:
ppt (g/kg) parts per thousand (grams/kilogram)
ppm (mg/kg) parts per million (milligrams/kilogram)
ppb (µg/kg) parts per billion (micrograms/kilogram)
% (g/100 g) percentage in weight (grams/100 grams)
mg/g milligrams/gram
mg/kg milligrams/kilogram
mol/kg moles/kilogram
mmol/g millimoles/gram
eq/kg equivalents/kilogram
meq/kg milliequivalents/kilogram