METHODS
5.5.10.2 Timed Increment
When
Timed Increment
is selected, the titrator acquires the potential (mV) at a fixed time
interval (no signal stability check).
The time period between two acquisitions must be set according to the reaction and the
response time of the electrode.
The range is from 2 to 180 seconds.