5-23
METHODS
If the stability condition is not reached and the maximum (t max) wait time has elapsed, the
potential is acquired.
delta E
- maximum change in potential during delta t
The range is from 0.1 to 99.9 mV.
delta t
- the time interval during which the potential is measured.
The range is from 0.5 to 10.0 seconds.
t min wait
- the minimum elapsed time before a stability check. This is also the minimum
elapsed time between two doses.
The range is from 2 seconds to
t max wait
time.
t max wait
- the maximum elapsed time between two successive doses. If the
t max wait
has
elapsed, a new dose is added even if the signal stability condition is not reached.
The range is from
t min wait
time to 180 seconds.