METHODS
5.5.10 Measurement Mode
During titration, the acquisition of the potential (mV) value of the solution can be done in
two ways: by using either
Signal Stability
or
Timed Increment
option.
5.5.10.1 Signal Stability
When
Signal Stability
is selected, the titrator acquires the potential (mV) only when stable
conditions are reached.
The principles of signal stability are plotted below:
The signal stability window (condition) represents the time interval (∆t) during which the
potential measured in solution (mV) is confined inside the potential interval (∆E).
The new signal value is acquired if the stability condition is reached after the minimum
(t min) wait time.