5-21
METHODS
To disable a pre-titration volume, enter 0.000 mL.
Note:
A pre-titration volume is highly recommended whenever possible. Fewer doses will
considerably shorten the overall titration duration.
5.5.9 Pre-Titration Stir Time
When enabled, the sample is mixed for a set period of time before any titrant is added. This
allows the sample to become homogeneous.
The range is from 0 to 180 seconds.
The
Pre-Titration Stir Time
option is disabled if 0 seconds is entered.