5-39
METHODS
5.5.18.1 Fixed Weight or Volume
Each titration will use a set weight or volume in the calculations.
5.5.18.2 Manual Weight or Volume
Each titration the exact weight or volume can be entered. The titrator will prompt for the
analyte weight volume at the beginning of each titration.
5.5.18.3 Same as previous (Linked Method Only)
The same weight or volume is used for both methods.
5.5.19 Titrant 1 Entry (Back Titration Only)
Select the mode for evaluating the necessary quantity of titrant 1 used in the back titration
process (phase 1).
5.5.19.1 Calculated by Formula
The volume of titrant 1 to be dispensed in the phase 1 of back titration will be calculated by
the titrator (see section
5.5.12.2 Back Titrations
).
5.5.19.2 Fixed by User
A fixed volume of titrant 1 will be used during the first phase of back titration process (direct
titration).