8-5
AUXILIARY FUNCTIONS
8.1.2 Rinse Tip
A 2 mL dose of titrant will be dispensed from the burette when this operation is selected.
This operation will eliminate the air from the dispensing tip.
8.1.3 Manual Dispense
Manual Dispense
option allows a defined titrant volume to be dosed. Select the
Manual
Dispense
option and press . The
Manual Volume Dispense
screen will become
active and the display will prompt you to enter the desired volume to be dispensed.
The manual dispense volume must be between the limits shown below:
0.001 to 4.500 mL for a 5 mL burette
0.001 to 9.000 mL for a 10 mL burette
0.005 to 22.500 mL for a 25 mL burette
8.1.4 Purge Burette
This option allows the burette to be emptied before cleaning and/or storing the burette. The
burette is flushed twice.
Note: Before starting this operation, remove the aspiration tube from the titrant bottle.