5-41
METHODS
5.5.22 Potential Range
The input potential range can be set by the user. The titration will be terminated and an error
message will appear if the potential is outside these limits.
These limits provide protection against a titration that does not generate an end point due to
potential over-range.
The ranges can be set between -2000.0 to 2000.0 mV.
5.5.23 Volume/Flow Rate
The flow rate for the dosing system can be set by the user in an interval of 0.1 to two times
the burette volume:
0.1 to 10 mL/min for a 5 mL burette
0.1 to 20 mL/min for a 10 mL burette
0.1 to 50 mL/min for a 25 mL burette
Note: The titrator will automatically detect the burette size, and display the correct high
limit volume.
The flow rate is set for all burette operations.