Method ID: HI0001EN
0.1N Sodium Hydroxide Titrant Concentration
Titrator Application Methods: General Methods - r. 2.3
Pre-Titration Stir Time: 60 sec
Measurement Mode: Signal Stability
delta E: 0.3 mV
delta t: 5 sec
t-min wait: 3 sec
t-max wait: 30 sec
Electrode Type: pH
Blank Option: No Blank
Calculations: Stdz. Titrant by Weight
Dilution Option: Disabled
Titrant Name: 0.1N NaOH
Analyte Size: 0.200 g
Analyte Entry: Manual
Maximum Titrant Volume: 15.000 mL
Stirring Speed: 1400 rpm
Potential Range: -2000.0 to 2000.0 mV
Volume/Flow Rate: 25 mL/50 mL/min
Signal Averaging: 1 Reading
Significant Figures: XXXXX
Calculations:
Calculations: Stdz. Titrant by Weight
Titrant units: N (eq/L)
Titrant volume dosed: V (L)
Standard weight: 0.200 g
(titrant/standard): 1.000 eq/mol
MW of standard: 204.23 g/mol




󰇛
󰇜
Results:
Titration Report
Method Name: 0.1N NaOH Titr. Conc.
Time & Date: 11:38 April 21, 2010
Titration ID: Ti_00010
Titration Results
Method Name: 0.1N NaOH Titr. Conc.
Time & Date: 11:38 April 21, 2010
Analyte size: 0.2182 g
End Point Volume: 10.056 mL
pH Equivalence Point: 8.273
Results: 0.10625 N (eq/L)
Initial and Final pH: 4.060 to 8.757
Titration Duration: 3:21 [mm:ss]
Titration went to Completion
Operator name: ____________________________
sale@ponpe.com
www.ponpe.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 902 Color and is the answer not in the manual?

Hanna Instruments HI 902 Color Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 902 Color
CategoryTest Equipment
LanguageEnglish

Related product manuals