Method ID: HI0010EN
0.1M Ferrous Ammonium Sulfate Titrant Concentration
(Ferrous Ammonium Sulfate – FAS)
Titrator Application Methods: General Methods - r. 2.3
Range: No
Filtered Derivatives: No
Pre-Titration Volume: 5.000 mL
Pre-Titration Stir Time: 0 sec
Measurement Mode: Signal Stability
delta E: 0.5 mV
delta t: 3.0 sec
t-min wait: 2 sec
t-max wait: 20 sec
Electrode Type: ORP
Blank Option: No Blank
Calculations: Stdz. Titrant by Weight
Dilution Option: Enabled
Final Dilution Volume: 100.000 mL
Aliquot Volume: 10.000 mL
Analyte size to be diluted: 0.490 g
Titrant Name: 0.1M FAS*6H2O
Analyte Size: 0.490 g
Analyte Entry: Manual
Maximum Titrant Volume: 15.000 mL
Stirring Speed: 1400 rpm
Potential Range: -2000.0 to 2000.0 mV
Volume/Flow Rate: 25 mL/50 mL/min
Signal Averaging: 1 Reading
Significant Figures: XXXXX
Calculations:
Calculations: Stdz. Titrant by Weight
Titrant units: M (mol/L)
Titrant volume dosed: V (L)
Standard weight: 0.490 g
Dilution factor: 0.10
Final dilution volume: 100.000 mL
Aliquot volume: 10.000 mL
(titrant/standard): 6.000 mol/mol
MW of standard: 294.18 g/mol
Results:
Titration Report
Method Name: 0.1M FAS Titr. Conc.
Time & Date: 13:31 April 22, 2010
Titration ID: Ti_00010
Titration Results
Method Name: 0.1M FAS Titr. Conc.
Time & Date: 13:31 April 22, 2010
Analyte size: 0.491 g
End Point Volume: 9.879 mL
mV Equivalence Point: 667.4
Results: 0.1024 M (mol/L)
Initial and Final mV: 791.3 to 598.0
Titration Duration: 3:05 [mm:ss]
Titration went to Completion