5-34
METHODS
A formula example is shown below using M (mol/L) as the titrant 1 units, M (mol/L) as the
titrant 2 units, mg/g and the final result units. This formula is used to calculate the amount of
titrant 1 to dispense:
The formula is based on the assumption that the sample concentration is 100% w/w.
The titrator will calculate the volume of titrant 1 needed to consume the sample and multiply
it with the excess factor in order to raise or lower the amount of titrant 1 dispensed.
Variables can be set according to the amount of sample and titrant used.
Press to proceed to the next formula.
If you do not want the titrator to calculate the volume of titrant 1 to add, see Section
5.5.18
Titrant 1 Entry
.
The remaining volume of titrant 1 needs to be calculated.
The following formula is used to calculate the remaining volume of titrant 1 after the reaction
with the sample: