IMPORTANT
The sample type and the sample type setting of the Element DC
must be same. Otherwise, the wrong test result will be obtained.
IMPORTANT
For further information of the reference interval settings, refer to
the Reference interval settings of MODE FUNCTION.
IMPORTANT
Be careful not to input wrong No. or ID.
10. Close the sampler cover.
11. Lock the sampler cover with the key.
WARNING
During sample processing, ensure the LCD under cover and sampler cover are closed to prevent injuries and
biohazard. When the display reads [Ready to Test] or [Testing], the sampler cover can be opened.
12. Start testing.
a. Touch START to begin testing.
IMPORTANT
The disposal box must remain closed during testing.
Otherwise, the results may be affected by opening it.
b. Automatic spotting starts. The spotting dialogue is
displayed.
NOTE: Touch STOP to cancel spotting. Those slides that have
already spotted will be measured, and test results will be
displayed.
c. [Spotting] on top of the dialogue will change to
[Testing] after spotting is complete, and the remaining
time to complete tests will be displayed.
NOTE: When performing the next test with the same No., ID,
Animal and Sample type as those of the previous test, touch
ADDITIONAL TEST. Only the setting of the Dilution field can be
changed.
NOTE: On basic dialogue, test preparation dialogue for the next
sample will be displayed as soon as spotting complete.
Displays estimated time (min.) to complete
the current measurements. When the
remaining time is less than 1 minute,
[Wait a minute.] is displayed.
34