A certain period of time has passed
Touch the panel
(Supplement) About Automatic Test Start
If the analyzer is not used for a period of time, the light source
lamp is turned off to save lamp life. This time period is set in the
lamp configuration MODE FUNCTION.
While the lamp is off, touch panel display goes dark.
By touching the panel to start tests again, the display indicates
[Lamp Warming Up].
d. After the test is completed, the test result dialogue is
displayed.
NOTE: When performing the next test with the same No., ID,
Animal and Sample type as those of the previous test, touch
the RERUN button. Only the setting of the Dilution field can be
changed.
NOTE: When performing a new test by changing the test factors,
touch the New test button.
13. Empty the disposal box when not performing a test. The
disposal box capacity is 100 slides and 50 tips.
Test result display area
WARNING
Do not open the disposal box during test processing. Otherwise, the analyzer may fail to properly eject used slides and
tips.
WARNING
Do not touch used slides and tips with you bare hands as this may increase your risk of infection. If you accidentally
touch those slides and tips, immediately wash your hands with running water and then use ethyl alcohol as a
disinfectant. Seek medical assistance if necessary.
IMPORTANT
Do not open the disposal box during test processing. Otherwise, adverse effects on test results may occur.
35