Fig. 102: ‘Program segment settings
Start date’
1 Start date
Fig. 103: ‘Program segment settings
Start date
Date selection’
1
[Conrmation]
button
2 Date selection
3
[Cancel]
button
Item Designation Description/function
7
Set temperature
‘Set tempera-
ture’ on page 65
You can change the temperature setpoint for this program
segment in steps of 0.1 with the
[Plus]
and
[Minus]
buttons.
By tapping on the temperature, a dialogue opens, via which
you can enter the temperature setpoint directly.
If a
‘-’
is given for the temperature setpoint, the incubator is
at a standstill during this time. A standstill segment is indi-
cated with a
‘-’
in the segment overview. A standstill seg-
ment makes sense for a program repetition.
Chapter 6.8.1 ‘Functional description of standstill mode’
on page 47
.
The temperatures can also be input outside the limits
of the temperature guard. This is not checked.
Please note that at the start of the program, the
input temperatures are compared with the tempera-
ture guard limits. If these were exceeded, a warning
message will be output
Chapter 6.13.5 ‘Tempera-
ture guard’ on page 98.
To input a start time (can only be set for the rst segment), proceed
as follows:
1. Tap on the start date.
The start time specied her is only for calcu-
lating the duration. The actual start time results
from the start of the created program or by the
activation of the timer
Further information
on page 84.
2. Tap on the desired date and conrm the selection with the
[Conrmation]
button.
You will go back to the
‘Program segment settings’
menu.