4. While monitoring the internal detector, the RF modulator level
is varied until the detected power is equivalent to the reference
power measured in step 2.
5. Modulation is re-enabled if appropriate.
These steps are performed in approximately 200 ps and are repeated
any time power or frequency is changed.
See Also
Softkeys listed above, Fltness Menu, (MOD), [POWER LEVEL),
Set
atten
“Externally Leveling the Swept Signal Generator” in Chapter 1
“Working with Mixers/Reverse Power Effects” in Chapter 1
“Working with Spectrum Analyzers/Reverse Power Effects” in
Chapter 1
Operating and Programming Reference A-9