SCPI
COMMAND SUMMARY
mapping of the frequency parameters on input to and output from
the swept signal generator. Changing this does not affect the
output frequency of the swept signal generator. Only the displayed
parameters and query responses are changed. The equation implied
by this is :
Entered/displayed frequency
=
(Hardware
Freq
* Multiplier ) + Offset
After
*RST,
the value is 0.
l
FREQuency:OFFSet:STATe
ONlOFFlllO
l
FREQuency:OFFSet:STATe?
Queries and turns the frequency offset off and on.
After
*RST,
the setting is OFF.
l
FREQuency:SPAN
<num>[freq
suffix]lMAXimumlMINimumlUPlDOWN
l
FREQuency:SPAN?
[MAXimumlMINimum~
Sets and queries the frequency span.
See FREQ : CENTER for more
information.
l
FREQuency:STARt
<num>[freq
suffix]
lMAXimumlMINimumlUPlDOWN
l
FREQuency:STARt?
CMAXimumlMINimuml
Sets and queries the START Frequency.
See FREQ : CENTER for more
information. *RST setting is MIN.
l
FREQuency:STEP:AUTO
ONlOFFlllO
l
FREQuency:STEP:AUTO?
Sets and queries the function switch that controls the calculation of
the frequency step
size (FREQ:STEP:INCR). If the response is
AUTO ON, then the swept signal generator selects a step size that is
coupled to the frequency span. The *RST setting is ON.
l
FREQuency:STEP[:INCRement]
<num>[freq
suffix]lMAXimumlMINimum
l
FREQuency:STEP[:INCRement]?
Sets and queries the frequency step size to use for any node in the
frequency subsystem that allows UP and DOWN as parameters. Setting
this value explicitly causes a
FREQ:STEP:AUTO OFF command.
The *RST setting is automatically calculated from span.
l
FREQuency:STOP
<num>Cfreq
suffixllMAXimumlMINimumlUPlDOWN
a
FREQuency:STOP?
[MAXimumlMINimuml
Sets and queries the stop Frequency. See
FREQ : CENTER for more
information.
*RST setting is MAX.
Operating and Programming Reference S-37