SCPI
COMMAND SUMMARY
Sets and queries the CW frequency. This does not change the
swept/CW
mode switch. *RST value is (MAX
+
MIN)/2
.
See FREQ : CENTER
for more information.
l
FREQuency[:CW]:AUTO
ONlOFFlllO
l
FREQuency[:FIXed]:AUTO
ONlOFFlllO
l
FREQuency[:CWl:AUTO?
l
FREQuency[:FIXed]:AUTO?
Sets and queries the
CW/
center frequency coupling switch. This
switch keeps
the two functions coupled together when ON. Changing
one
ofthem,
changes both.
*RST setting is OFF. See FREQ:CENTER
for more information.
l
FREQuency:MANual
<num>[freq
suffix]lMAXimumlMINimumlUPlDOWN
l
FREQuency:MANual? [MAXimumlMINimuml
Sets and queries the
manual frequency. This controls the output
frequency in swept manual mode. The limits
are START and STOP.
*RST
valueis
the same as FREQ:CENTER. See FREQ:CENTERfor more
information.
l
FREQuency:MODE
FIXedlCWlSWEeplLIST
l
FREQuency:MODE?
Sets and queries the switch that selects either swept, CW or list
operation. *RST value is CW.
l
FREQuency:MULTiplier
<num>lMAXimumlMINimum
l
FREQuency:MULTiplier? [MAXimumlMINimuml
Sets and queries the frequency multiplier. The numeric value is
rounded to the nearest integer. This function changes mapping of
frequency parameters on input to and output from the swept signal
generator. Changing this does not affect the output frequency of the
swept signal generator. Only the displayed parameters and query
responses are changed.
The equation implied by this is :
Entered/displayed frequency
=
(Hardware
Freq
*
Multiplier ) + Offset
After
*RST,
the value is 1.
l
FREQuency:MULTiplier:STATe ONlOFFlllO
l
FREQuency:MULTiplier:STATe?
Queries and turns the frequency multiplier off and on.
After
*RST,
the setting is OFF.
l
FREQuency:OFFSet
Cnum>lMAXimumlMINimum
l
FREQuency:OFFSet? [MAXimumlMINimuml
Sets and queries the frequency offset. This function changes the
S-36 Operating and Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8360 B Series and is the answer not in the manual?

HP 8360 B Series Specifications

General IconGeneral
BrandHP
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals