(SWEEP
TIME)
Function Group
Menu Map
Description
Programming Codes
See Also
SWEEP
7
This
hardkey
lets you set a sweep time for frequency sweeps or power
sweeps. The sweep time range is 10 ms to
2OOs,
but the fastest sweep
time is constrained by the frequency span. The fastest possible sweep
can be determined automatically:
1. Press SWEEP
(MENU),
this reveals the sweep menu keys.
2. Select more
l/3
, to scroll to the next page of the sweep menu.
3.
Select
SwpTime
Auto,
to set the sweep time to automatic. The
swept signal generator calculates the fastest possible calibrated
sweep time for any sweep span.
Whenever you press
(SWEEP),
the active entry area displays
the current sweep time and whether the sweep time is coupled to
the frequency span (far right hand side displays: AUTO). If the word
AUTO is not displayed then the sweep time auto function is off.
SCPI: SWEep[:FREQ
uency]:TIME <num>[time suffix] or
MAXimumIMINimum
Analyzer: ST
<num>
[sclms]
Power Sweep
“Power Level and Sweep Time Operation” in Chapter 1
“Programming Typical Measurements” in Chapter 1
SwpTime
Auto
Function Group
SWEEP
Menu Map
7