EasyManua.ls Logo

Impulse Dymanics OPTIMIZER Smart System - Show OMNI II Programmer Client Window; Starting a Remote Session with the OMNI Smart Programmer

Impulse Dymanics OPTIMIZER Smart System
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
87
Note: To use the remote operation feature of the OMNI Smart Programmer, a
user must receive a unique user name and password from Impulse Dynamics. An
appropriate Client Name must also be registered with Impulse Dynamics.
Once all of the above information has been entered, press Enter or click on the
Connect button. The status of the connection will appear in the white space on
the lower part of the window (Status). If the connection is successful, the OMNI
II Programmer Client window will disappear (the remote session started). If an
error occurs (incorrect password or connectivity issues for instance) the OMNI II
Programmer Client window will remain open, and the appropriate error message
will appear in the Status space, requiring the user to retry the connection.
A remote linkup can then be requested from an OMNI Smart Programmer in
Remote mode:
A Clinical Engineer at the Remote Follow-up Center (or any other
location, as long as the Clinical Engineer has reliable internet access and
telephone connection) who has a Remote OMNI Smart Programmer
connected to the Internet can initiate pairing with the Clinical OMNI
Smart Programmer at the requesting clinic.
If internet connectivity is interrupted during the remote session, the Local
(Slave) OMNI Smart Programmer will display a warning screen, and
CCM delivery by the OPTIMIZER Smart IPG being followed-up will be
turned ‘OFF’. The Local (Slave) OMNI Smart Programmer will
automatically revert to Master mode after 5 consecutive failed retry
attempts on any communication order.
The local (Slave) OMNI Smart Programmer will also revert to Master
mode after pressing the ‘Urgent’ Programming button on the OMNI II
Programmer Wand.
31.4.2 Show OMNI II Programmer Client Window
In order to show the OMNI II Programmer Client window after the Programmer
has logged on to the Remote Server:
Select the Remote button on the Task Bar.
Select the Client button on the Tool Bar.
31.5 Starting a Remote Session with the OMNI Smart Programmer
To start a remote session with the OMNI Smart Programmer, the clinician at the clinic
has to first call the Remote Follow-up Center via the telephone to request a remote
follow-up session. Then the clinician has to connect their OMNI Smart Programmer to
the Internet via an Ethernet or wireless network connection.

Table of Contents