68
For the OPTIMIZER Smart IPG the log browser window is composed of the following
elements:
Time scale that can be modified with the available options or typing a scale value.
Idx button that opens up a Marker index log window containing log items that can
be used for navigation purposes.
Markers and ECG labeled display.
Horizontal scrollbar to allow browsing in time.
Vertical scrollbar whenever the signals do not fit in the defined window.
Print button.
When an event in the Marker index log window is double-clicked, the Marker Window
screen automatically displays the information associated to the event time stamp.
Analogous behavior occurs when a command in the log window or the associated
indicator in the screen is selected.
The distance between two available cursors, one principal to be fixed and another to be
moved along the window, allow direct measurements of several dimensions as time
intervals or voltage.
24. CCM Schedule
When the OPTIMIZER Smart IPG is programmed to deliver Timed CCM therapy, the CCM
Schedule parameters become enabled.
To access the CCM Schedule parameters:
Select the Parameters button on the Task Bar
Select the CCM Train &Schedule tab on the Parameters bar.
The upper middle half of this tab contains the parameters Start Time and End Time. The
parameters set the general start and end time of CCM therapy delivery during each day. By
default, the CCM therapy schedule is set to be distributed over a 24 hour period each day. The
standard programming for these parameters is:
Start Time: 0 h 0 m
End Time: 23 h 59 m
The upper right half of this tab contains the parameters On Time and Off time. The On Time
parameter sets the amount of time the OPTIMIZER Smart IPG is scheduled to deliver CCM
therapy within the time frame set by the Start time and End time. The Off Time parameter sets
the amount of time the OPTIMIZER Smart IPG is scheduled not to deliver CCM therapy within
the time frame set by the Start time and End time.
The upper left half of this tab is the parameter Scheduled. The Scheduled parameter sets the
total number of hours per day the OPTIMIZER Smart IPG is scheduled to deliver CCM therapy
within the time frame set by the Start time and End time. This parameter is calculated
depending on settings for On Time and Off time.