26.1 Loading Statistics
To obtain the Statistics from the OPTIMIZER Smart IPG:
• Place (or replace, if necessary) the Programmer Wand over the OPTIMIZER
Smart IPG implant site.
• Select the Statistics button on the Task Bar.
• Select the Read button at the bottom of the Statistics bar.
If the loading is successful, the Programmer will display the message “Read Statistics
OK”.
However if the Programmer Wand is not well positioned over the implant site, the
transfer of data may fail. If a failure in communication occurs, the Programmer will
display the message “Read Statistics Error” along with the options Retry and Cancel.
Retry can be performed by repositioning the Programmer Wand and selecting the Retry
button.
When the OPTIMIZER Smart IPG is in Marker Mode and within communications range
of the OMNI II Programmer Wand, the contents of the Statistics counters change
dynamically, showing the current number of occurrences of each event type.
26.1.1 Statistics Tabs
• On – General
o Events
Atrial: number of atrial events sensed during scheduled CCM
delivery
Ventricular: number of ventricular events sensed during
scheduled CCM delivery
LS in Alert: number of local sense events sensed during scheduled
CCM delivery
o Periods
Normal: number of normal periods during scheduled CCM
delivery
Inhibited: number of inhibited periods during scheduled CCM
delivery
Post-Inhibited: number of post-inhibited periods during scheduled
CCM delivery
o Trains Delivered
During Onset: number of CCM trains delivered during the onset
of scheduled CCM delivery
Total: total number of CCM trains delivered during scheduled
CCM delivery