114
Electromagnetic interference information (Table 5 of 6):
Recommended separation distances between portable and mobile RF communications equipment and the
OMNI Smart Programmer or OPTIMIZER Mini Charger
Both the OMNI Smart Programmer and the OPTIMIZER Mini Charger should be used in an electromagnetic
environment with limited radiated RF noise. The customer or user of the OMNI Smart Programmer or
OPTIMIZER Mini Charger can help prevent electromagnetic interference by maintaining the minimum distance
between portable and mobile RF communications equipment (transmitters) and the OMNI Smart Programmer or
OPTIMIZER Mini Charger recommended below, which is determined by the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance broken down by transmitter frequency
(m)
150 kHz to 80 MHz
1
__
d = 1.17 √ P
80 MHz to 800 MHz
1
__
d = 0.35 √ P
800 MHz to 2.5 GHz
__
d = 0.70 √ P
0.01
0.12
0.04
0.07
0.1
0.37
0.11
0.22
1
1.17
0.35
0.7
10
3.7
1.11
2.22
100
11.7
3.5
7.0
For transmitters with a maximum rated output power not listed above, the recommended separation distance d in meters (m)
can be estimated by using the equation applicable to the frequency of the transmitter, where P is the maximum rated output
power of the transmitter in watts (W) specified by the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply to all settings. Electromagnetic propagation is affected by absorption and reflection
from buildings, objects, and people.
Electromagnetic interference information (Table 6 of 6):
Compliance with the Radio and Telecommunications Terminal Equipment (R&TTE) Directive
(1999/5/EC)
The OMNI Smart Programmer complies with the Radio and Telecommunications Terminal Equipment (R&TTE)
Directive (1999/5/EC).
Applicable standard
Compliance requirements
ETSI EN 302 195-1: 2004
ETSI EN 302 195-2: 2004
EN 60601-1-2
1
Radiated H Field (normal and extreme conditions)
Spurious emissions (radiated), (transmitter)
Frequency range of the modulation (normal and extreme conditions)
Spurious Radiation (receiver)
1
Compliance is claimed to EN 60601-1-2 rather than the two harmonized standards (ETSI EN 301 489-1 and ETSI EN 301
489-31). This is due to the fact that not all sections of the ETSI standards are applicable and those that are applicable are
covered by testing performed in EN 60601-1-2 which uses equivalent test methods and performance criteria which are as
restrictive as or even more restrictive than those in ETSI EN 301 489-1 and ETSI EN 301 489-31.