106
Factory Settings
PARAMETERS RELATED TO CONTROLLING RIGHT HEART SENSING
Mode OOO
Atrial Sense Amplifier Sensitivity
1.3 mV
Ventricular Sense Amplifier Sensitivity 2.0 mV
Ventricular Sensing Polarity Bipolar
Atrial Sensing Polarity Bipolar
Ventricular Refractory Period
250 ms
Post-Ventricular Atrial Refractory Period 250 ms
CCM PULSE TRAIN ACTIVATION
CCM Pulse train enable OFF
CCM PULSE TRAIN TIMING
Number of pulses 2
Train delay
35 ms
Phase 1 duration 5.14 ms
Phase 2 duration 5.14 ms
Phase 1 polarity Positive
Phase 2 polarity Negative
CCM Pulse Amplitude 7.5 V
CCM signal delivery channel
LS, V
Interval 0 ms
CCM INHIBIT ALGORITHM
CCM Inhibit Count
2 beats
Short AV Delay
70 ms
Long AV Delay
398 ms
Atrial tachycardia rate
154 bpm