10. OMNI SMART PROGRAMMER SYSTEM
10.1 Description
The OMNI Smart Programmer System allows the physician to interrogate and program
the OPTIMIZER Smart IPG. The programmer software runs on a Lenovo Touch screen
Laptop connected to a Programmer Interface box. Communication between the
Programmer Interface and the OPTIMIZER Smart IPG is accomplished with a
Programmer Wand placed directly over the implant site. The Programmer Wand
communicates via magnetic induction telemetry with the OPTIMIZER Smart IPG
implanted in the patient. Understanding the instructions contained in this manual on how
to operate the OMNI Smart Programmer System is essential to proper operation of an
OPTIMIZER Smart IPG.
Figure 3: OMNI Smart Programmer and Programmer Wand
Warning: The OMNI Smart Programmer may be subject to interference from other
electrical devices operated in the vicinity. Portable and mobile RF equipment
is especially likely to impair the normal function of the programmer. If the
OMNI Smart Programmer is not operating as expected, such interference has
to be always taken into account. Other equipment, even if it outperforms the
CISPR emission limits, can also interfere with the OMNI Smart Programmer.